Date: |
2021-03-15 |
From: |
Factory - none |
To: |
Lab - none |
Title: |
Returned |
Notes: |
Returned from Eosense for high temp start problem and noisy CO2 readings. It had a failed clock crystal causing start up issues at high temperature, it was given custom firmware to bypass the component. It was calibrated for CO2 concentration and CO2 flux |
Entered by: |
Joe Verfaillie |
|
|