Equipment and Calibration History

View Equipment |

View history entry 5198 for Forerunner-Eosense eosFD sn: FD20180047

Date: 2021-03-15
From: Factory - none
To: Lab - none
Title: Returned
Notes: Returned from Eosense for high temp start problem and noisy CO2 readings. It had a failed clock crystal causing start up issues at high temperature, it was given custom firmware to bypass the component. It was calibrated for CO2 concentration and CO2 flux
Entered by: Joe Verfaillie