View history entry 5142 for Licor LI-7500A sn: 75H-2514

Date: 2021-03-01
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2020-12-11 AGC=101.1 SDM=0 DAC1=- DAC2=- Ambient P=100.4kPa, Tair=27.2C, Tdew=0.92C Zero Air at 500psi, AMP CA04396 408.97 at 250 Tdewspan=20.0 CO2 zero 0.9049 -> 0.9048 Offset: -0.73ppm CO2 span 1.0082 -> 0.9984 Offset: 5.03ppm H2O zero 0.8398 -> 0.8389 Offset: -6.7ppm H2O span 1.0093 -> 1.0061 Offset: 0.07ppm Notes: Flow rates were larger than usual - 2lpm instead of 2cfpm
Entered by: