View history entry 5106 for Gill WM 1590 sn: 153805

Date: 2021-02-01
From: Lab - none
To: Lab - none
Title: Tested
Notes: After being removed from a partially submerged position on the fallen BC tower, I tested this sensor in the lab. There was no evidence that water had leaked inside. I connected it to a Licor system and used fans to blow air through in on the floor. The response sign and magnitude looked about right. I plan to run it on the Tonzi tower for a couple weeks to check its performance.
Entered by: Joe Verfaillie