Date: |
2021-02-01 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Tested |
Notes: |
On the 29th I removed this sensor from being submerged after the BC tower fell just after midnight on the morning of the 27th. On the evening of the 29th, about 1/2 cup of water came out of the bottom when I removed the base. I removed the desiccant bottle in the head and saw no evidence that water had leaked in there. I left dry air flowing inside the base over night. On the 30th I removed the bottom case and only saw minor water damage on the pressure sensor. On the 1st, after consultation with Licor, I powered up the analyzer and got green power and status lights but no communications on the Ethernet. It will need to be returned to Licor for further evaluation. |
Entered by: |
Joe Verfaillie |
|
|