View history entry 5103 for Licor LI-7700 sn: TG1-0155

Date: 2021-02-01
From: Lab - none
To: Lab - none
Title: Tested
Notes: On the 29th I removed this sensor from being submerged after the BC tower fell just after midnight on the morning of the 27th. On the evening of the 29th, about 1/2 cup of water came out of the bottom when I removed the base. I removed the desiccant bottle in the head and saw no evidence that water had leaked in there. I left dry air flowing inside the base over night. On the 30th I removed the bottom case and only saw minor water damage on the pressure sensor. On the 1st, after consultation with Licor, I powered up the analyzer and got green power and status lights but no communications on the Ethernet. It will need to be returned to Licor for further evaluation.
Entered by: Joe Verfaillie