Equipment and Calibration History

View Equipment |

View history entry 4996 for Licor LI-7500A sn: 75H-2176

Date: 2020-12-02
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2020-08-25 AGC=101 SDM= DAC1= DAC2= Ambient P=100.8kPa, Tair=25.1C, Tdew=3.3C Zero Air at 450psi, AMP CA4396 408.97 at 300 Tdewspan=20.00 CO2 zero 0.8671 -> .8673 Offset: .59ppm CO2 span 1.0055 -> 1.009 Offset: -1.94ppm H2O zero 0.8526 -> .8531 Offset: 3.0ppm H2O span 0.9246 -> .9238 Offset: .02ppm Notes:
Entered by: