View history entry 4995 for Licor LI-7500A sn: 0042

Date: 2020-12-02
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2020-08-25 AGC=101 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=24.3C, Tdew=2.2C Zero Air at 450psi, AMP CA4396 408.97 at 300 Tdewspan=20.00 CO2 zero 0.9479 -> .9482 Offset: 1.16ppm CO2 span 1.0073 -> 1.0111 Offset: -2.03ppm H2O zero 0.8441 -> .8436 Offset: -3.1ppm H2O span 1.0268 -> 1.0264 Offset: .01ppm Notes: 2nd time doing calibration today. The first calibration had large offsets, so I decided to redo it (CO2 zero offset 3.93, CO2 span offset -3.77, H2O zero offset -13.5, H2O span offset 0.39). The 2nd time looks much better with expected range of offsets, but I'm not sure why it was different.