Date: |
2020-11-30 |
From: |
Lab - |
To: |
Lab - |
Title: |
Calibration |
Notes: |
Calibration Location: Lab
Last calibration: 2020-05-20
RSSI=73%, Optical RH=11%
Ambient P=101.2kPa, Tair=24.0C
Zero Air at 500psi, AMP CA4396 1.9497 at 375
CH4 Zero Offset: -0.054ppm
CH4 Span Offset: 0.009ppm
Notes: First calibration after renaming from "unknown" to its serial number. After this calibration, I will upgrade the firmware to v1.0.29.
|
Entered by: |
|
|
|