View history entry 4983 for Licor LI-7700 sn: TG1-0155

Date: 2020-11-20
From: Bouldin Corn - Gas Analyzer CH4
To: Bouldin Corn - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: Bouldin Corn Last calibration: 2020-05-19 RSSI=88%, Optical RH=20% Ambient P=102.5kPa, Tair=9.4C Zero Air at 575psi, AMP CA3923 3.121 at 550 CH4 Zero Offset: 0ppm CH4 Span Offset: 0.009ppm Notes: Foggy day with a litle breeze. The RSSI after cleaning was 50, but jumped up to 88 after putting hood on and running dry air through.