Equipment and Calibration History

View Equipment |

View history entry 4944 for Licor LI-7700 sn: TG1-0424

Date: 2020-11-12
From: Sherman Wetland - Gas Analyzer CH4
To: Sherman Wetland - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: Sherman Wetland Last calibration: 2020-05-21 RSSI=82%, Optical RH=12% Ambient P=101.8kPa, Tair=14.7C Zero Air at 700psi, AMP CA3923 3.121 at 700 CH4 Zero Offset: -0.084ppm CH4 Span Offset: 0.002ppm Notes: Still day
Entered by:
Field notes