Equipment and Calibration History

View Equipment |

View history entry 4890 for Licor LI-7500A sn: 0042

Date: 2000-10-12
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Zero with N2 was 2.0, after running several weeks in the lab.Cal gas 402.25 read 411.1 ppm at P=101.31 kPa and T 23.17 C. Licor Dewpoint sensor read 14.57 Tdew. Vaisala standard read 13.9 C. We are using the Vaisala as the standard for the experiment. Rezeroed and spanned D/A #1, H2O mmol m-3 0v=0 5 v=1000 D/A #2, CO2 mmol m-3 0v=13 mmol m-3 5v=20 mmol m-3
Entered by: Joe Verfaillie