View history entry 4671 for Licor LI-7500A sn: 75H-2150

Date: 2020-08-20
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2020-02-11 AGC=99.8 SDM= DAC1= DAC2= Ambient P=100.1kPa, Tair=27.3C, Tdew=14.5C Zero Air at 725psi, AMP CA4396 408.97 at 500 Tdewspan=20.00 CO2 zero 0.903 -> .9019 Offset: -4.84ppm CO2 span 1.0154 -> 1.0014 Offset: 7.21ppm H2O zero 0.8674 -> .8665 Offset: -5.7ppm H2O span 1.0224 -> 1.0110 Offset: 0.26ppm Notes: Zero offsets and CO2 span larger than I expected. I had trouble with the H2O span at first because the water concentration kept decreasing. I added DI water to the dewpoint generator reservoir, which fixed the issue. Air quality was bad inside from wildfire smoke leaking in through windows.