Date: |
2020-08-20 |
From: |
Factory - none |
To: |
Lab - none |
Title: |
Received |
Notes: |
Report from CSI: Sensor communicates ok and draws normal current, it does read OOT so it will need a new cal and the rotted 20' cable replaced.
Repair: Cleaned up the sensor element, updated OS, calibrated the sensor and had a new 20'cable made. |
Entered by: |
Daphne Szutu |
|
|