Equipment and Calibration History
View Equipment |
Date: |
2020-07-23 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Tested |
Notes: |
This sensor browned out the power at East End. I found one bad nick in the cable but after opening it up I did not see any damage to the conductors there. I plugged it into a power supply in the lab and it immediately pulled the 12V supply down to 7V. This needs to be sent to Campbell. |
Entered by: |
Joe Verfaillie |
|
|