Equipment and Calibration History

View Equipment |

View history entry 4524 for Apogee S2-431 sn: 1061

Date: 2020-06-11
From: Lab - none
To: Lab - none
Title: Tested
Notes: The two red/far red sensors stopped working on the BA tower. I tested them on the CR1000 in the lab running the BA program and still got no response. I tried the CR1000 terminal to talk to them over SDI-12 and got nothing trying both C1 and C3 ports. I will get a USB to SDI-12 from Apogee soon and will try again with that.
Entered by: Joe Verfaillie