Date: |
2011-11-18 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Calibrated in the lab with god bottle 1.8978ppm CH4 at 180psi. Last calibration Aug 12, 2011. Before running the cal gas the LGR read:
10:33PST, 21.58C, 142.54Torr, 15.466usec, 1Hz, 1.866ppm
I missed writing down the ppm with the cal gas before calibration but it was about 1.80ppm. Afterwards it was right on. |
Entered by: |
|
|
|