Equipment and Calibration History

View Equipment |

View history entry 4499 for Campbell Sci. CS547A sn: 7157

Date: 2020-05-26
From: Sherman Wetland - Conductivity Sensor
To: Sherman Wetland - Conductivity Sensor
Title: Calibration
Notes: Cleaned and calibrated in the field with DI water (0 mS) and a YSI solution (10,000 uS). These are the temperature-corrected mScm readings from the datalogger, where the temperature correction occurs in the datalogger program. In DI water: 0.56 mScm In 10,000 uS calibration soln: 10.6 mScm Calibration: (Corrected mScm) = 0.996 * (Sensor mScm) - 0.558
Entered by: Daphne Szutu