Equipment and Calibration History

View Equipment |

View history entry 4486 for Licor LI-7700 sn: TG1-0223

Date: 2020-05-21
From: Mayberry - Gas Analyzer CH4
To: Mayberry - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: Mayberry Last calibration: 2019-10-22 RSSI=83%, Optical RH=26% Ambient P=101.2kPa, Tair=21.8C Zero Air at 750psi, AMP CA3923 3.121 at 800 CH4 Zero Offset: 0.004ppm CH4 Span Offset: -0.003ppm Notes: Breezy day. Swapped optics drierite after calibration and used cal gas to dry head space. New optics RH was 9%.
Entered by:
Field notes