Equipment and Calibration History

View Equipment |

View history entry 4483 for Licor LI-7700 sn: TG1-0357

Date: 2020-05-21
From: Sherman Barn - Gas Analyzer CH4
To: Sherman Barn - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: Sherman Barn Last calibration: 2019-10-31 RSSI=80%, Optical RH=12% Ambient P=101.6kPa, Tair=19.2C Zero Air at 800psi, AMP CA3923 3.121 at 800 CH4 Zero Offset: 0.036ppm CH4 Span Offset: 0.036ppm Notes: Breezy day
Entered by:
Field notes