Equipment and Calibration History

View Equipment |

View history entry 446 for Cole Parmer 32707-30 sn: CP6295

Date: 2011-10-20
From: Factory -
To: Factory -
Title: Calibration
Notes: Initial calibration. Flow rates at STP, max allowable error +-1.5%, Cal for Air, at 23C, 740mmHg, 12.8VDC ml/min = 202.56 * VDC - 8.8158, R2 = 0.9998 ml/min = 1.0115 * screen - 7.7203, R2 = 0.9998 Cal sheet on file.
Entered by: