Equipment and Calibration History

View Equipment |

View history entry 4449 for Licor LI-7500A sn: 75H-2176

Date: 2020-05-13
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2019-11-13 AGC=100 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=21.6C, Tdew=11.9C Zero Air at 725psi, AMP CA4396 408.97 at 550 Tdewspan=16.00 CO2 zero 0.8675 -> .8672 Offset: -1.42ppm CO2 span 1.0072 -> 1.0081 Offset: -.52ppm H2O zero 0.852 -> .8521 Offset: .8ppm H2O span 0.9123 -> .9159 Offset: -.09ppm Notes:
Entered by: