View history entry 4448 for Licor LI-7500A sn: 75H-2667

Date: 2020-05-13
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2019-11-12 AGC=102 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=21.3C, Tdew=10.6C Zero Air at 725psi, AMP CA4396 408.97 at 500 Tdewspan=16.00 CO2 zero 0.9419 -> .9422 Offset: 1.29ppm CO2 span 1.019 -> 1.0196 Offset: -0.25ppm H2O zero 0.8384 -> 0.8380 Offset: -2.3ppm H2O span 1.0239 -> 1.0313 Offset: -0.15ppm Notes: