Equipment and Calibration History

View Equipment |

View history entry 4447 for Licor LI-7500A sn: 0042

Date: 2020-05-13
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2019-12-04 AGC=99.9 SDM= DAC1= DAC2= Ambient P=100.9kPa, Tair=19.9C, Tdew=9.18C Zero Air at 750psi, AMP CA4396 408.97 at 550 Tdewspan=16.00 CO2 zero 0.949 -> .9481 Offset: -3.44ppm CO2 span 1.0182 -> 1.0141 Offset: 2.03ppm H2O zero 0.8441 -> .8441 Offset: 0.1ppm H2O span 1.0341 -> 1.0271 Offset: 0.15ppm Notes:
Entered by: