Equipment and Calibration History

View Equipment |

View history entry 4371 for Licor LI-7500A sn: 0073

Date: 2020-02-11
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2019-10-30 AGC=100.3 SDM=0 DAC1=- DAC2=- Ambient P=100.8kPa, Tair=26.4C, Tdew=-2.8C Zero Air at 800psi, AMP CA04396 408.97 at 550 Tdewspan=16.0 CO2 zero 0.9574 -> 0.9572 Offset: -0.8ppm CO2 span 1.0091 -> 1.0096 Offset: -0.37ppm H2O zero 0.7311 -> 0.7313 Offset: 0.9ppm H2O span 1.0052 -> 1.0091 Offset: -0.08ppm Notes: Ambient air very dry.
Entered by: