View history entry 4360 for Licor LI-7500A sn: 75H-2150

Date: 2020-02-04
From: Lab -
To: Lab -
Title: Maintenance
Notes: I started to calibrate this IRGA but the zeros were way high: ~58ppm CO2 and ~60umol/m3 H2O. The span seemed okay and other IRGAs calibrated well. I think the internal chemicals are shot. I replaced the chemicals - they were hard and I had to scrape them out with a screw driver. After putting in fresh chemicals I flushed the body with zero air before sealing it up. The zeros immediately fell by about halt to ~23ppm CO2 and ~30umol/m3. I left the IRGA running and will try to calibrate late tomorrow for field installation the next day.
Entered by: Joe Verfaillie