View history entry 4230 for Licor LI-7700 sn: TG1-0155

Date: 2019-12-04
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Calibration Location: Lab Last calibration: 2019-10-22 RSSI=87%, Optical RH=7% Ambient P=99.9kPa, Tair=20.2C Zero Air at 900psi, AMP CA4396 1.9497 at 650 CH4 Zero Offset: -0.091ppm CH4 Span Offset: 0.011ppm Notes: Testing this sensor after it fell down with the BC tower. First calibration. Offsets seem reasonable. May be a little less stable , or I was just impatient waiting for the span gas to equilibrate.