View history entry 4190 for Vaisala HMP60 sn: L1740398

Date: 2019-10-15
From: Lab - none
To: Lab - none
Title: Tested
Notes: Tested at Oxford Tract for a week against reference sensor HMP155 sn R0360206. See RT Report "HMP155 lab and Oxford test": https://nature.berkeley.edu/biometlab/bmetdata/rtreport.php?screen=view&id=28 The sensor was within spec, so no calibration needed. (Specifications from manual: AirT= +/-0.6°C, RH= +/-3% for 0-90% RH)
Entered by: Daphne Szutu