View history entry 4151 for Licor LI-7500A sn: 75H-2176

Date: 2019-11-13
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2019-07-24 AGC=99.5 SDM= DAC1= DAC2= Ambient P=100.5kPa, Tair=25.6C, Tdew=11.4C Zero Air at 900psi, AMP CA4396 408.97 at 650 Tdewspan=16.00 CO2 zero 0.8672 -> .8675 Offset: 1.36ppm CO2 span 1.0061 -> 1.0072 Offset: -.67ppm H2O zero 0.8521 -> .8520 Offset: -1.0ppm H2O span 0.9182 -> .9123 Offset: .13ppm Notes: Did an initial calibration yesterday using config file 2019-07-24, but it had a large H2O zero offset (-35.7 mmol/m3) and H2O span offset (1.53C). Today I did another calibration, again from the config file 2019-07-24. Today the H2O offset/span seem ok, so not sure what was happening yesterday. I saved today's new config file and deleted yesterday's to avoid confusion.