Equipment and Calibration History

View Equipment |

View history entry 4126 for Licor LI-7500A sn: 0073

Date: 2019-10-30
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2019-08-14 AGC=99.6 SDM=0 DAC1=- DAC2=- Ambient P=101.0kPa, Tair=27.45C, Tdew=-10.8C Zero Air at 900psi, AMP CA04396 408.97 at 800 Tdewspan=20.0 CO2 zero 0.9576 -> 0.9574 Offset: -0.70ppm CO2 span 1.0095 -> 1.0091 Offset: 0.23ppm H2O zero 0.7318 -> 0.7311 Offset: -4.6ppm H2O span 1.0247 -> 1.0052 Offset: 0.4ppm Notes: Ambient air was very dry. Joe
Entered by: