Date: |
2019-10-22 |
From: |
Bouldin Alfalfa - Gas Analyzer CH4 |
To: |
Bouldin Alfalfa - Gas Analyzer CH4 |
Title: |
Calibration |
Notes: |
Calibration Location: Bouldin Alfalfa
Last calibration: 2019-04-11
RSSI=84%, Optical RH=22%
Ambient P=101.8kPa, Tair=20.9C
Zero Air at 1300psi, AMP CA03923 3.121 at 1400
CH4 Zero Offset: 0.022ppm
CH4 Span Offset: 0.001ppm
Notes: Swapped optics desiccant after calibration using cal gas to dry head space. New RH 7%
|
Entered by: |
|
|
|