Equipment and Calibration History

View Equipment |

View history entry 4113 for Licor LI-7700 sn: TG1-0155

Date: 2019-10-22
From: Bouldin Corn - Gas Analyzer CH4
To: Bouldin Corn - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: Bouldin Corn Last calibration: 2019-04-11 RSSI=89.5%, Optical RH=26% Ambient P=102.1kPa, Tair=17.2C Zero Air at 1350psi, AMP CA03923 3.121 at 1400 CH4 Zero Offset: 0.024ppm CH4 Span Offset: -0.005ppm Notes: Light breeze, swapped optical desiccant after calibration using cal gas to dry head space. New RH 7%.
Entered by:
Field notes