Calibration |
Notes: |
Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 06/2018" for details. I calculated a calibration factor of 12.7 uV/(W/m2), but more testing needs to be done because there was a large calibration difference when testing the sensor upside-down (15.4 uV/(W/m2)). |