Equipment and Calibration History

View Equipment |

View history entry 4046 for Licor LI-7500A sn: 0073

Date: 2019-08-14
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2019-04-23 AGC=98.7 SDM= DAC1= DAC2= Ambient P=100.2kPa, Tair=29.6C, Tdew=13.2C Zero Air at 1000psi, AMP CA4396 408.97 at 850 Tdewspan=16.00 CO2 zero 0.9563 -> .9576 Offset: 4.5ppm CO2 span 1.0094 -> 1.0095 Offset: -.13ppm H2O zero 0.7336 -> .7318 Offset: -10.5ppm H2O span 1.0153 -> 1.0247 Offset: -.5ppm Notes: When the SS was 98.7, I reset it to 100. The previous calibration on 2019-04-23, the H2O zero was offset by 12.1 mmol/m3, and this time the offset was -10.5 mmol/m3.
Entered by: