Date: |
2019-04-25 |
From: |
Mayberry - Gas Analyzer CH4 |
To: |
Mayberry - Gas Analyzer CH4 |
Title: |
Calibration |
Notes: |
Calibration Location: MB
Last calibration: 2018-10-26
RSSI=80%, Optical RH=12%
Ambient P=100.9kPa, Tair=26C
Zero Air at 1600psi, AMP CA3923 3.121 at 1600
CH4 Zero Offset: 0.013ppm
CH4 Span Offset: 0.015ppm
Notes: TC module swapped before calibration; breezy day
|
Entered by: |
|
|
|