View history entry 3930 for Licor LI-7500A sn: 0073

Date: 2019-04-23
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2019-01-07 AGC=98.6 SDM= DAC1= DAC2= Ambient P=100.4kPa, Tair=26.0C, Tdew=11.8C Zero Air at 1100psi, AMP Air 408.97 at 700 Tdewspan=16.00 CO2 zero 0.956 -> .9563 Offset: .91ppm CO2 span 1.0081 -> 1.0094 Offset: -0.78ppm H2O zero 0.7316 -> .7336 Offset: 12.1ppm H2O span 1.0023 -> 1.0153 Offset: -.26ppm Notes: