Equipment and Calibration History

View Equipment |

View history entry 3899 for Licor LI-7700 sn: TG1-0423

Date: 2019-04-11
From: East End - Gas Analyzer CH4
To: East End - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: East End Last calibration: 2018-10-30 RSSI=73%, Optical RH=11% Ambient P=101.9kPa, Tair=16.5C Zero Air M4002 at 1700psi, AMP CA03923 3.121 at 1800 CH4 Zero Offset: 0.036ppm CH4 Span Offset: 0.013ppm Notes: Windy day
Entered by:
Field notes