Date: |
2019-04-11 |
From: |
East End - Gas Analyzer CH4 |
To: |
East End - Gas Analyzer CH4 |
Title: |
Calibration |
Notes: |
Calibration Location: East End
Last calibration: 2018-10-30
RSSI=73%, Optical RH=11%
Ambient P=101.9kPa, Tair=16.5C
Zero Air M4002 at 1700psi, AMP CA03923 3.121 at 1800
CH4 Zero Offset: 0.036ppm
CH4 Span Offset: 0.013ppm
Notes: Windy day
|
Entered by: |
|
|
|