Equipment and Calibration History
View Equipment |
Date: |
2019-02-12 |
From: |
Sherman Wetland - Gas Analyzer CH4 |
To: |
Sherman Wetland - Gas Analyzer CH4 |
Title: |
Upgrade |
Notes: |
Upgraded firmware (embedded software) from 1.0.23 to 1.0.27. Licor change log: "Expanded a buffer in the DSP to prevent phase changes in the DSP data caused when the buffer filled up. This fixes the synchronization loss and resultant data loss between the LI-7550 and LI-7700." |
Entered by: |
Daphne Szutu |
|
|
Field notes