Equipment and Calibration History

View Equipment |

View history entry 3847 for Vaisala HMP45A sn: U2120013

Date: 2019-01-24
From: Mayberry - RH & Temp
To: Mayberry - RH & Temp
Title: Serviced
Notes: Short in the 12V wire at the splice point. Repaired in the field with a screw terminal, but should be brought back to the lab for a full repair.
Entered by: Daphne Szutu
Field notes