View history entry 3837 for Licor LI-7500A sn: 0073

Date: 2019-01-07
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2018-09-26 AGC=100 SDM=0 DAC1=0 DAC2=0 Ambient P=100.9kPa, Tair=18.7C, Tdew=9.5C Zero Air at 1200psi, AMP Air 408.97 at 850 Tdewspan=16.00 CO2 zero 0.9569 -> .9560 Offset: -2.67ppm CO2 span 1.0112 -> 1.0081 Offset: 1.65ppm H2O zero 0.7314 -> .7316 Offset: .6ppm H2O span 1.0183 -> 1.0023 Offset: .35ppm Notes:
Entered by: