Equipment and Calibration History

View Equipment |

View history entry 3761 for Licor LI-7700 sn: TG1-0155

Date: 2018-10-25
From: Bouldin Corn - Gas Analyzer CH4
To: Bouldin Corn - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: Bouldin Corn Last calibration: 2018-04-04 RSSI=91%, Optical RH=16% Ambient P=101.5kPa, Tair=27.2C Zero Air at 1925psi, AMP Air 3.121 at 1975 CH4 Zero Offset: .015ppm CH4 Span Offset: -.013ppm Notes: Gently breezy day
Entered by:
Field notes