Date: |
2018-10-25 |
From: |
Mayberry - Gas Analyzer CH4 |
To: |
Mayberry - Gas Analyzer CH4 |
Title: |
Calibration |
Notes: |
Calibration Location: MB
Last calibration: 2018-04-25
RSSI=81%, Optical RH=8%
Ambient P=101.6kPa, Tair=20.5C
Zero Air at 2000psi, AMP Air 3.121 at 2100
CH4 Zero Offset: 0.021ppm
CH4 Span Offset: 0.018ppm
Notes: Calm day
|
Entered by: |
|
|
|