Date: |
2018-10-02 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Last calibration: 2018-09-26
AGC=101 SDM=0 DAC1=0 DAC2=0
Ambient P=99.9kPa, Tair=26.8C, Tdew=15.9C
Zero Air at 1600psi, AMP Air 408.97 at 1200 Tdewspan=20.00
CO2 zero 0.9392 -> .9389 Offset: -1.47ppm
CO2 span 1.0153 -> 1.0172 Offset: -1.09ppm
H2O zero 0.9056 -> .9072 Offset: 9.2ppm
H2O span 1.0166 -> 1.0238 Offset: -.16ppm
Notes: First calibration after swapping out internal chemicals. Kept the sensor on for 3 days after the swap before the calibration.
|
Entered by: |
|
|
|