Equipment and Calibration History

View Equipment |

View history entry 3566 for Licor LI-7500A sn: 75H-2176

Date: 2018-07-10
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2018-03-05 AGC=101.3 SDM= DAC1= DAC2= Ambient P=100.5kPa, Tair=27C, Tdew=13.3C Zero Air at 175psi, LBL Air 408.97 at 1400 Tdewspan=16.00 CO2 zero 0.8679 -> .8671 Offset: -3.65ppm CO2 span 1.0057 -> 1.0029 Offset: 1.23ppm H2O zero 0.8486 -> .848 Offset: -3.4ppm H2O span 0.9112 -> .9088 Offset: .05ppm Notes:
Entered by: