View history entry 3565 for Licor LI-7500A sn: 75H-2150

Date: 2018-07-10
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2018-03-05 AGC=97 SDM= DAC1= DAC2= Ambient P=100.5kPa, Tair=27.2C, Tdew=12.7C Zero Air at 150psi, LBL Air 408.97 at 1300 Tdewspan=16.00 CO2 zero 0.9023 -> .9017 Offset: -2.64ppm CO2 span 1.0093 -> 1.002 Offset: 3.79ppm H2O zero 0.9655 -> .8662 Offset: 5.1ppm H2O span 1.0096 -> 1.0069 Offset: .06ppm Notes:
Entered by: