View history entry 3563 for Licor LI-7500A sn: 0041

Date: 2018-07-03
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2018-03-05 AGC=00 SDM=0 DAC1=0 DAC2=0 Ambient P=100.1kPa, Tair=22.2C, Tdew=12.5C Zero Air at 300psi, LBL Air 408.97 at 1300 Tdewspan=16.00 CO2 zero 0.9972 -> .9966 Offset: -1.94ppm CO2 span 1.0196 -> 1.0173 Offset: 2.0ppm H2O zero 0.8194 -> .8192 Offset: -.9ppm H2O span 1.0267 -> 1.0285 Offset: -.03ppm Notes: Training Robert
Entered by: