Equipment and Calibration History

View Equipment |

View history entry 3562 for Licor LI-7700 sn: TG1-0357

Date: 2018-07-10
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Calibration Location: Lab Last calibration: 2017-11-09 RSSI=74.6%, Optical RH=2% Ambient P=100.5kPa, Tair=25.8C Zero Air at 220psi, LBL Air 1.9497 at 1400 CH4 Zero Offset: -0.001ppm CH4 Span Offset: 0.0127ppm Notes: