Date: |
2011-05-13 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Maintenance |
Notes: |
On first power up CH4 and H2O did not line lock. I adjusted the VO setting in the linelock_0.ini file from -0.690 to -0.590.
I did a calibration with the CH4 god bottle 1.8978ppm CH4, 350psi. Last cal was:
CH4 - 14 Jul 2010
H2O, CO2 - 13 Mar 2009
Precal: 14:26PST, 22.77C, 138.11Torr, 14.052usec, 1Hz, 2.058ppm
After cal: 1.9021ppm |
Entered by: |
|
|
|