Equipment and Calibration History

View Equipment |

View history entry 3537 for Licor LI-7500A sn: 75H-2667

Date: 2018-06-25
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-12-15 AGC=105.0 SDM=0 DAC1=- DAC2=- Ambient P=100.4kPa, Tair=25.6C, Tdew=12.2C Zero Air at 350psi, AMP 408.97 at 1350 Tdewspan=15.99 CO2 zero 0.9415 -> 0.9414 Offset: -0.3ppm CO2 span 1.0207 -> 1.0162 Offset: 2.28ppm H2O zero 0.8372 -> 0.8375 Offset: 1.8ppm H2O span 1.0166 -> 1.0237 Offset: -0.15ppm Notes: Calibrated by Joe
Entered by: