Equipment and Calibration History

View Equipment |

View history entry 3399 for Licor LI-7500A sn: 75H-2514

Date: 2018-04-18
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-12-15 AGC=102.5 SDM=0 DAC1=0 DAC2=0 Ambient P=101.0kPa, Tair=22.2C, Tdew=2.5C Zero Air at 500psi, LBL Air 408.97 at 1400 Tdewspan=16.00 CO2 zero 0.9052 -> .9060 Offset: 3.56ppm CO2 span 1.0118 -> 1.0106 Offset: 0.64ppm H2O zero 0.8385 -> .8368 Offset: -11.4ppm H2O span 1.0103 -> 1.0160 Offset: -0.12ppm Notes:
Entered by: