Date: |
2018-04-13 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Calibration Location: Lab
Last calibration: 2017-10-11
RSSI=75%, Optical RH=4%
Ambient P=101.8kPa, Tair=24.2C
Zero Air at 500psi, LBL Air 1.9497 at 1500
CH4 Zero Offset: .008ppm
CH4 Span Offset: -.0417ppm
Notes: Started "new" ultra pure zero air tank that had 500 psi left in it.
|
Entered by: |
|
|
|