View history entry 3394 for Licor LI-7500A sn: 0073

Date: 2018-04-13
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-11-27 AGC=99 SDM=0 DAC1=0 DAC2=0 Ambient P=101.9kPa, Tair=23.7C, Tdew=2.9C Zero Air at 100psi, LBL Air 408.97 at 1500 Tdewspan=16.00 CO2 zero 0.9571 -> .9575 Offset: .99ppm CO2 span 1.0118 -> 1.0145 Offset: -1.26ppm H2O zero 0.7306 -> .7295 Offset: -6.9ppm H2O span 1 -> 1.0131 Offset: -.3ppm Notes: